Stopping a Run
page 51
Sally Sue and Peggy Sue User Guide
Stopping a Run
1.
To stop a run, click
Stop
. When the run stops, instrument status will change to not ready and a cleanup
button displays:
NOTE: If a run is stopped before it completes, Sally Sue and Peggy Sue must perform a cleaning protocol to
remove used capillaries from system trays. This prepares them for the next run.
2.
Click
Cleanup
.
Let Sally Sue or Peggy Sue finish the cleaning protocol, which takes about ten minutes. Once cleaning is
done, instrument status will change to ready and you can start a new run.
Waste and Water Bottles
You should empty the waste bottle and refill the water bottle in the Auxiliary Module before starting a new
run. To remove or install bottles:
IMPORTANT
Do not remove either bottle if Sally Sue or Peggy Sue are running, doing this may abort the run in
progress.
1.
Remove the bottle by unscrewing it clockwise until it disengages from the manifold.
2.
Pull the bottle down slowly until the sensor at the bottom of the bottle is clear.
3.
Fill the water bottle to the base of its neck and empty the waste bottle as needed.
4.
To reinstall the bottle, lift it up to the manifold making sure to enclose the sensor that sits at the bottom
inside the bottle. Screw the bottle counter-clockwise until the arrow on the bottle's alignment label
lines up with the arrow on the manifold's alignment label.
Summary of Contents for ProteinSimple Peggy Sue
Page 12: ...page 8 Chapter 2 Getting Your Lab Ready Sally Sue and Peggy Sue User Guide ...
Page 36: ...page 32 Chapter 3 Sally Sue and Peggy Sue Sally Sue and Peggy Sue User Guide ...
Page 64: ...page 60 Chapter 4 Operating Sally Sue and Peggy Sue Sally Sue and Peggy Sue User Guide ...
Page 76: ...page 72 Chapter 5 Maintenance and Troubleshooting Sally Sue and Peggy Sue User Guide ...