AllInOneCycler
TM
/
Thermal Block
60
www.bioneer.com
BQ-042-101-01
Revision : 0(2012-12-07)
Procedure
(1)
Turn off the instrument and disconnect the power cord from the back of the
All
I
nOneCycler
TM
.
ELECTRICAL SHOCK HAZARD
Severe electrical shock, which could cause physical injury or death, can result from working
on an instrument when the high voltage power supply is operating. To avoid electrical shock,
disconnect the power supply to the instrument, unplug the power cord, and wait at least 1
minute before working on the instrument.
(2)
Insert a small screw-driver (-) into the slot in the upper portion of the power entry module,
and open the fuses door.
Figure 53. The power entry module in the back side of the
All
I
nOneCycler
TM
(3)
Pull out the fuse compartment to inspect the two fuses in the fuse compartment.
Figure 54. Two fuses in the fuse compartment
(4)
Pull out the blown fuse(s) from the fuse compartment and replace with new 250V, F10AL, 5×
20mm fuses.