AllInOneCycler
TM
/
Thermal Block
24
www.bioneer.com
BQ-042-101-01
Revision : 0(2012-12-07)
Table 6. How to set incubate options.
Option
Description
Extended time per
cycle
This menu is to set the gradual time increase and
decrease of the selected step with each cycle.
1.
Touch and change the values between 0 and 59.
2.
Touch
button behind the value to select increase or
decrease of time.
3.
When this option is applied, the set value will be
displayed on the step of the procedure graph with this
figure
.
Increment
temperature per
cycle
This menu is to set the gradual temperature increase
and decrease of the selected step with each cycle.
1.
Touch and change the values between 0 and 2.0.
2.
Touch
button behind the value to select increase or
decrease of temperature.
3.
When this option is applied, the set value will be
displayed on the step of the procedure graph with this
figure
.
Ramping Rate
This menu is to set the constant value of the ramping
rate.
1.
Touch and change the values between 0.1 and 6.0.
2.
When this option is applied, the set value will be
displayed on the step of the procedure graph with this
figure
.