Chapter 7 Shielding Installation
68
Source and center of the detector have to be installed on a hori-
zontal line. At the same time, this horizontal line is the monitoring
level where the device switches.
The shielding is installed at the measuring point by means of
brackets.
1
Install suitable bracket which has to be provided by the cus-
tomer on the respective level. See installation proposal below.
2
Unpack source with shielding and mount it on the respective
bracket.
If you are working with a shielding with pneumatic shutter, a com-
pressed air connection has to be available. Signals supplied by an
optional limit switch have to be connected via an electrical line
(see Shielding LB 744X with Pneumatic on page 132).
7.2
Installation Proposal for Shielding
The shielding container comprises a lead-filled sturdy cast iron
case. A revolving shutter is installed to close the radiation exit
channel. This shutter is operated from the rear side via a lever,
which can be locked by means of a padlock in its open or closed
position.
For installation, the shielding container includes a cast-on flange
and in addition a fastening foot with threaded holes.
The specifications for the required drilled holes are listed in the
Appendix on page 131.
Arrangement
Figure 33:
Installation on
a container
Installation sequence
Shielding
Monitoring level
Bracket provided
by customer
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...