Chapter 4 Getting Started
56
4.4
Calibration in Standard Mode
Empty calibration has to be performed in order to calibrate the
measurement.
Prerequisites (see also page
Container must be empty or at least below the monitoring
level.
The gas density existing under actual operating conditions is
available in the container. If this is not possible, select the pro-
fessional mode and compensate the gas density value with
code 35. The special features of the professional mode have to
be taken into account.
A possible existing heating or cooling jacket is filled with the
medium to be used later.
In agitator containers, the agitator must in operation, if it may
have any influence on the limit value.
The shielding with the source has been installed. The radiation
exit channel is open.
The EVU is not in the edit mode.
Calibration:
Push Cal button for 3 seconds.
The seconds are counted down in the top row.
The bottom row shows the count rate.
In code 20 you can read off the measured zero count rate.
The measurement is now calibrated and already sup-
plies live measured values.
Note!
A system check has to be carried out to ensure that the measure-
ment operates correctly.
To do this, fill the container above fill level limit, or alternatively,
close the shielding when the container is empty.
59s
24
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...