Chapter 3 Parameters
26
Note!
Leave the time constant on automatic unless there are good rea-
sons for entering a fixed value.
If you enter a fixed value as time constant, you should keep in
mind that a fixed value does not adjust to the source decay.
Therefore, enter a value that ensures a safe switching function
even after 10 years.
Rule of thumb: automatically calculated time constant x 10.
If the container is under high gas pressure or if significant wall
caking is to be expected, then it is absolutely essential that you
enter a fixed value. The value should be so great that possible gas
density variations or wall caking cannot trigger the alarm.
Rule of thumb: automatically calculated time constant x 30.
Live count rate (cps)
Display of non-average count rate.
(Only in the professional mode)
The displayed count rate comes directly from the detector and is
not averaged by the time constant.
Maximum time constant (s)
Upper limit of time constant in code 12
(Only in the professional mode)
Input
Maximum time constant
0
Disabled
> Code 12
Enabled
If the time constant in code 12 is set to automatic, then the time
constant in code 12 is automatically increased or adapted to the
source decay, so that no false alarm can be triggered by statistical
fluctuations. To avoid that the time constant may become too big
over the years, you can here enter a value that triggers an alarm
as soon as the time constant in code 12 exceeds this value. The
display shows the error message 39.01 "Replace source". Code 55
shows the year in which the message is likely to be output.
Please keep in mind: If a fixed value has been entered in code 55,
then the error message "Replace source" is displayed as soon as
this date is exceeded. Code 14 is then disabled.
C o d e 1 3
C o d e 1 4
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...