Chapter 3 Parameters
25
Reaction of the alarm relay in case of error
Define reaction of alarm relay in case of error
(Only in the professional mode)
Input
Reaction of alarm relay in case of error
0
Alarm relay holds the last status
1
Alarm relay follows the error relay
You have to define if the alarm relay should follow the error relay
in case of error.
Reading (%)
Live reading in % level
The reading is dependent on the count rate in code 11 and the
measuring range in code 20 and 21, corresponding to 0 to 100%.
The averaging of the reading is dependent on the time constant in
code 12.
To be able to jump quickly to code 10 it suffices if you keep the
Clear button pushed down for 2 seconds. Prerequisite is that you
are not in the edit mode and that code 10 is the standard result
display. The standard result display is set in code 15.
Reading (cps)
Live reading in counts per second
The averaging of this count rate is dependent on the time constant
in code 12.
Time constant (s)
Time constant for reading averaging
(Only in the professional mode)
Input
Time constant on:
Code
-01.0
Automatic
C
>0
Fixed value
C
If you enter -01.0 (factory setting), the time constant is aver-
aged automatically. The time constant is continuously adapted to
the source decay and is set optimally.
For further explanations see page 66.
A fixed value will be accepted only if the input is equal or greater
than the automatically determined time constant.
The dot in the displayed result is a decimal point which separates
seconds from tenths of a second.
C o d e 0 9
C o d e 1 0
C o d e 1 1
C o d e 1 2
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...