Empty calibration
The level must be at least 50 mm below the limit level. The
radiation channel of the shielding is open.
Push the “Cal” button for 3 seconds
The empty count rate is read in.
At the end of the measuring time, code 30 is displayed and
the read-in empty count rate.
Full calibration
The level must be at least 50 mm above the limit level.
If this is not possible, you may also close the source shield-
ing. If only the source shielding is closed to carry out the
full calibration, then you have to keep in mind:
-
Make sure that the absorption of the closed shielding
nearly corresponds to the absorption of the product. If in
doubt, please contact B
ERTHOLD
T
ECHNOLOGIES
or your local
representative.
-
Typically, you will lose some dynamics in the measure-
ment, which results in a higher time constant.
Select code 31.
Push the Enter button
Push the Cal button.
The full count rate is read in.
At the end of the measuring time, code 31 is displayed and
the read-in full count rate.
Calibration
Select code 36.
Push the Cal button.
A checkmark and the digits 0000 confirms that the calibra-
tion was successful.
If an error message is display, the error has to be removed
as described in the User's Manual and in the error list; then
the calibration has to be carried out once more.
Test
Write down the empty and full count rate in code 30 and
code 31.
Select code 60.
Enter the full count rate in code 60.
The measurement must show 0% in code 10.
If not, check the calibration.
Select code 60.
Enter the full count rate in code 60.
The measurement must show 100% in code 10.
If not, check the calibration.
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...