Chapter 3 Parameters
27
Standard reading display
Define standard reading display
(Only in the professional mode)
Input
Standard reading
10
Code 10
11
Code 11
Here you can define to which reading the display will jump if no
input has been made for 20 minutes.
Minimum / Maximum Limit Switch
Define mode of operation of the limit switch
Input
Alarm relay function
0
Max. alarm switch
1
Min. alarm switch
Here you can define if the alarm relay should switch as max. or
min. alarm.
If set correctly, the relay drops out and the contact between ter-
minal 12a and 12c is open if an alarm is triggered.
Max. alarm switch:
If an alarm is to be triggered in case of overfilling.
Min. alarm switch:
If an alarm is to be triggered in case of underfilling.
Switching threshold (%)
Switching threshold in percent
(Only in the professional mode)
Input
Limit value
Code
-1
Automatic
C
>0
Fixed value in %
C
If the switching threshold is exceeded, the device indicates an
alarm.
If the switching threshold is set to automatic (factory setting), it is
already set up optimally. If you wish to set another switching
threshold, you can enter a fixed value in % level.
You can enter the switching threshold either as a percentage value
in this code or in code 18. Every entry in this code automatically
calculates the new count rate in code 18. Vice versa: If you enter a
fixed value (count rate) in code 18, then the percentage value in
code 17 is calculated new.
C o d e 1 5
C o d e 1 6
C o d e 1 7
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...