Chapter 6 Error Messages
69
Error
number
Error
sub-code
Designation
Display
Reaction of
measurement
Alarming relay
Cause
Remedy
35
01
1
ECR < ZCR
2
!!
RUN
-
Empty count rate lower than
zero count rate.
Determine empty count rate or /
and zero count rate new.
02
ECR < FCR
3
Empty count rate lower than full
count rate.
Determine empty count rate or /
and full count rate new.
03
FCR < ZCR
Full count rate lower than zero
count rate.
Determine full count rate or / and
zero count rate new.
04
C 12 > C 14
Time constant is higher than the
maximum time constant in code
14
Check time constant.
Check calibration data or
increase code 14.
05
Distance switch
point to FCR
Distance switch point to full
count rate is too small. Danger
of switching errors.
Adjust switch point, or increase
time constant.
06
Distance switch
point to ECR
Distance switch point to full
count rate is too small. Danger
of switching errors.
Adjust switch point, or increase
time constant.
07
Hysteresis
too large
Hysteresis too large.
Increase time constant or reduce
hysteresis.
08
Time constant too
small
Time constant too small
Increase time constant or set switch
point further into the center.
09
Time constant
too big
Time constant exceeds the
maximum value of 999s.
Check the calibration and the
parameter settings.
39
01
Source replace-
ment
!!
RUN
Dependent on
code 44
The source activity is so low that
the source has to be replaced
within the next 12 months. The
error has been triggered by code
14 or code 55.
Check year for source replacement
in code 55!
Check time constant in code 12
and 14!
Check calibration values in code 20,
21!
If necessary, replace source.
40
01
Safe switching
function not
ensured
!!
RUN
Dependent on
code 44
Switch point does not have the
required distance to empty or
full calibration
or time constant is too low.
Check calibration. Adjust switch
point setting or/and time constant.
Table 8: Error messages
1
ECR = empty count rate
2
ZCR = zero count rate
3
FCR = full count rate
Chec
ked onl
y
in cas
e of
a
cal
ibra
ti
on
in code 3
6
.
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...