57
Restore the mains power to the
NOxBOX
i
.
If it is not possible to restore the
mains power, ensure the
Manual Override is prepared to
switch-on. See Section 6.12.
The system will automatically
shut-down when the battery charge
runs out.
No vent flow detected by the
NOxFlow for more than 30 seconds.
Ensure the NOxFlow is installed
correctly in the patient inspiratory
limb. The green printed arrow points
towards the patient.
This message may also appear
where insufficient ventilator flow is
present, such as with very low
tidal volume applications.
Increasing the bias flow may help
resolve this alarm.
The NOxBOX
i
has detected a critical
fault within the intelligent delivery
system, and can no longer
guarantee safe delivery function.
Engage the Manual Override (see
Section 6.12) and replace the
system as soon as practical.
Contact the Systems Engineer.
6.9.2 Gas Supply Alarms
An amber banner warning alarm
is triggered when one cylinder is
running low whilst there is a good
second supply cylinder detected.
To resolve this alarm, switch off the
empty gas cylinder at the valve.
Remove the gas supply line from
the rear of the monitor and
depressurise the line. The system
will automatically switch to the full
cylinder for continuous treatment.
Alternatively press the cross
button to mute the alarm for 2
minutes and discuss the banner
message.
Alarms