61
Zero Calibration is automatically performed upon system power-on. The results of the zero
calibration are displayed during the set-up procedure prior to performing the system test.
During use, the system automatically reminds you once every 24 hours to perform the zero
calibration procedure. This helps maintain the accuracy of the monitoring system and is a regular
quick check to ensure safe performance of the system.
You can perform the zero calibration procedure at any time. To perform a zero calibration from the
main screen:
1.
Select the Settings tab,
press Sensor Zero.
The zero test will take
approximately 2 minutes to
run. A progress animation
is displayed under the test
table.
6.
If all sensors perform within
the expected parameters, a
“Zeroing Successful”
message is displayed, with
green check marks against
each sensor result.
5.
2.
When entering the zero
calibration screens from either
the alarm state flag or from
the settings screen, the
following message is
displayed.
3.
Press Next to continue with
the zero calibration.
4.
Press Back to return to the
previous screen.
7.
Press Next to return to the
main screen. The monitored
gas values will take a minute
to read the true gas delivery
values.