34
16.
If the system test failed, follow the diagnostic tips to correct:
a.
Ensure that all connections on the Test Line Assembly are made as directed.
b.
Ensure the NOxFlow sensor is orientated with the air flow in the direction of the
arrow marking, the dose portion closest to the sample end of the test line.
c.
Ensure the valves are open on all NO supply cylinders.
d.
Check the air flow is set to 10 L/min.
17.
If the system test fails a second time, the unit cannot be used for intelligent automated
delivery. Call the System Engineer, and replace the system.
18.
Once the system test has passed; shut off the O
2
flow, and disconnect the tested Line
Assembly from the supply line. The verified Line Assembly is ready to connect into the
patient ventilator circuit
If the NOxBOX
i
system is not to be used within next 10 minutes, turn
off the cylinder valves, detach the supply lines from Port 1 and Port
2, and release the line pressure using the purge needle at the back of
the monitor. Then re-attach the supply lines to Port 1 and Port 2.
WARNING: If the NOxBOX
i
system is left pressurized for more than 10
minutes while not in use, repeat the system test. Connect the Line
Assembly to the air supply as described in steps 11-17. Press Next.
The system test will run automatically.
Ventilator inspiratory limb showing NOxFlow placement before humidifier.
19. Connect the NOxBOX
i
to the patient ventilator inspiratory limb. See Section 5.3 onwards for
Patient Circuit connection diagrams.
The sample point adaptor should be connected immediately behind the patient Y-piece on
the inspiratory limb. To achieve consistent results, place a one-way flutter valve between the
sample connector and the Y-piece to prevent contamination of the sample from exhaled
breath.
20 . Connect the upstream end of the NOxFlow sensor to the upstream inspiratory limb tubing
The NOxFlow should be situated upstream of the humidifier.
Maintain the optimum 1m distance of ventilator tube between the sample point and the dose
introduction point on the NOxFlow for best performance (between 0.7 and 1.3m is ideal).
The system is now ready for dose entry and alarm setting procedures.