62
8.
If one or more sensors fail
to perform within the
expected parameters
during the zero calibration,
a “Zeroing Failed”
message will appear, and
any sensors that failed to
meet the performance
criteria will display a red
cross next to them.
9.
Press Next for instructions.
10.
If the zero calibration fails,
the system will prompt to
contact the service
engineer and replace the
system as soon as
possible. The system will
still return to its normal
monitoring display and
can be used to deliver NO,
but caution must be used
when operating the system
with an active sensor fault.
6.11 User Settings Screen
Changing Display Language
On the User Settings Screen,
change the system display
language by selecting the
appropriate name