If the dose entered is greater than
20 ppm, a warning will appear to
alert you that a high clinical dose
value has been entered.
If a dose value of zero is
entered, the system will alert user
that this will stop the
treatment.
NOTE:
All alarms will still function unless
the regulator hoses are
diconnected , where the system
will enter stand by mode.
3.
Press next to continue
4.
The system will display a
confirmation screen for the
dose entered.
5.
To accept the displayed
value, press the
check mark.
6.
To change this value,
press the Edit
The system displays a summary
of the entered dose and calculated
alarm values.
•
The system calculates a
default alarm value for the
NO High Alarm based on
the confirmed NO dose
entry.
•
The system calculates a
default alarm setting for the
NO Low Alarm based on the
confirmed NO dose entry.
•
The system has a default
NO
2
High Alarm setting of
1 ppm.
•
The system has a default
O
2
Low Alarm setting of
19 %.