53
6.7.3 Change Dose and Alarms
When changing the dose, the system recalculates the alarm values to
appropriate levels for the new dose setting. These values can be
confirmed or edited before returning to the Main Screen.
To change the dose from the main
screen, press the dose edit button.
The system will display the direct entry
dose keypad. Follow the system
instructions the same as when setting
the initial dose.
To change the gas alarm settings, press
the button corresponding to the alarm to
edit (image shows edit high NO alarm).
The edit dose screen shows the current
alarm value. Press the delete arrow to
clear this value and input a new value.
Press next to proceed, the system will
return to the main screen once the new
setting is accepted.
NOTE: the alarm settings have limits
beyond which it will not allow the alarm
to be set.