66
The system screen will display a
warning message that the system is
delivering through the Manual Override
mode.
Press the cross to dismiss this
message.
The “NO Dose” area of the main screen
shows a constant reminder that the
system is on Manual Override.
To revert to intelligent delivery, position
the mode selection valve to vertical.
6.12.4 Manual Override Flow Selector: Approximate NO flow values
50
50 cc/min
100
100 cc/min
200
200 cc/min
300
300 cc/min
400
400 cc/min
500
500 cc/min
Note all values shown are approximate to +/-
15%. Refer to look-up tables for indication of
dose per ventilator flow rate.
Monitoring of patient dose must be
implemented when using Manual
Override flow delivery of NO.
6.13 Shutdown
When ready to shutdown the system, press and
hold the power button on the monitor for two
seconds until the System Shutdown screen is
displayed.
Confirm shutdown on screen. Followon-screen
instructions.
600
600 cc/min