65
4.
Turn Manual override manual flow control valve
to the value that most closely matches the desired
dose in the event of engaging the Manual
Override.
5.
Once flow is set, the system is ready to allow safe
engagement of the Manual Override mode if
required during this treatment.
6.12.2 Using the Printed Look-up Tables
From the NOxBOX
i
printed Operating
Instructions, refer to the lookup tables.
Select the table for the NO cylinder
concentration being used
Find the ventilator minute volume flow
that most closely matches the patient
ventilator settings.
Each flow setting from 50-600 ml displays
the approximate expected NO dose
delivery for this minute volume flow
based on a 1M length of tubing between
theNOxFLOW and the sample line
connection.
Turn the flow valve to the setting most
closely matched to the desired patient
dose.
6.12.3 Engage Manual Override:
Ensure the Manual Override Flow Valve
is positioned to the most appropriate
flow setting (see right).
Turn mode selection valve to the
manual override position. The system
will now deliver a continuous stream of
NO into the ventilator stream, ensuring
continuous treatment until a
replacement system is ready.