7
Failure Identification Code Troubleshooting Table
7-12
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
4
5
Upstream occlusion was detected.
Check for spilled solution in
sensor region. Ensure set is proper
type and is properly loaded.
Remove upstream occlusion and
reposition tubing. Clean the sensor
with cotton swabs dampened with
one of the agents listed in Table
6-1. Remove fibers or foreign
particles. Do not use hard
instruments for cleaning.
Ensure ambient and solution
temperatures are between 60° and
100° F (15.5° and 37.7° C).
Raise or lower temperatures.
Perform the appropriate steps
Calibration Mode 1” on page 9-14
related to upstream occlusion.
Reading displayed in pump’s PRI
RATE window must be within
3242 and 3314. If not, recalibrate
or replace the sensor.
Check for damage to door latch
and hinges.
Replace any damaged parts.
6
Low battery voltage (10.4 V or
less) was detected.
When BATTERY LOW alarm is
on and CHARGING LED is off,
check:
a. Tightness of AC plug
b. AC voltage source
c. Power fuse
d. Battery fuse
a. Plug in firmly.
b. When AC voltage is below 105
VAC, connect pump to a correct
voltage source.
c. Replace when it has failed.
d. Replace when it has failed.
Turn pump off and recharge for
16 hours.
When alarm recurs after
recharging, check and repair loose
connections at battery terminals. If
problem persists, replace battery.
7
Communication alarm was
detected: A framing error, overrun
error or parity error occurs during
the communication with an
external computer.
Check that the pump is in
computer control mode. Check for
the loss of the communications
link. Verify that computer
software is correctly loaded onto
PC.
Place pump into computer control
mode.
Ensure communication cable is
plugged in and/or the cable is
wired correctly.
Correct problems with PC
software.
8
Communication time out alarm
was detected: No communication
between the pump and an external
PC for a given period
See Failure Identification Code 7.
Place pump into computer control
mode.
Ensure communication cable is
plugged in and/or the cable is
wired correctly.
Correct problems with PC
software.
Table 7-3
Failure Identification Codes — continued
Code
Cause
Checks
Corrective Action
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...