07-19-B1-729
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
7-3
Troubleshooting
Determing the Software Version
Determing the Software Version
In order to troubleshoot the pump you must know the pump’s installed
software version. Perform the following to detemine the pump’s software
version:
1.
Turn the pump off.
2.
Press and hold the
SILENCE
key, then press the
ON-OFF/CHARGE
key.
3.
Record the software version displayed in the secondary VTBI display.
Alarms
The pump has a number of built-in safety features. Should a situation occur
which requires operator attention or intervention, the pump stops infusing
and sounds an audible alarm. The following are brief descriptions of these
alarms.
AIR:
An ultrasonic sensor in the pump head detects air in the administration
set. Detection of an air bubble stops the infusion and illuminates the red
ALARM LED. AIR is displayed in the Message Display and the audible
alarm is activated.
OCCLUSION:
The downstream occlusion sensor senses an increase in
back-pressure between the patient and the pump, indicating an occlusion or
closed clamp. When an occlusion is sensed, the pump stops, OCCLUSION
is displayed in the Message Display, the red ALARM LED is illuminated
and the audible alarm is activated.
UPSTREAM OCCLUSION:
The upstream occlusion sensor senses a
closed clamp or complete blockage upstream of the pump. When sensed, the
pump stops, UPSTREAM OCCLUSION is displayed in the Message
Display, the red ALARM LED is illuminated, and the audible alarm is
activated.
BATTERY LOW:
When approximately 15 minutes of running time remain
during battery operation, BATTERY LOW is displayed in the Message
Display, the yellow ALERT LED is illuminated, and the audible alert is
activated. After approximately 15 minutes, the pump stops, BATTERY
LOW is displayed on the Message Display, the red ALARM LED is
illuminated, and the audible alarm is activated.
Table 7-1 the alarms and the possible causes of each. In all cases, fluid
infusion is halted. Appropriate troubleshooting is discussed throughout the
rest of this chapter.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...