10
Exterior Inspection
10-2
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
Note:
Prior to beginning these tests, please become familiar with the pump
operating procedures in the Operator’s Manual.
Exterior Inspection
1.
Inspect the exterior of the pump.
Q
Verify that the pump is clean and free from physical damage.
Q
Check the gray RTV surrounding the front keypad. Ensure an intact,
uniform bead of sealant is present. Perform touch up as required and
allow to cure.
2.
Check the mounting clamp.
Q
Verify the clamp screw rotates smoothly without binding.
Q
Verify that the disc and cushion are intact and secure.
3.
Check the power cord.
Q
Verify that there are no cuts or breaks that would expose an electrical
shock hazard.
Q
Check the strain relief and plug. Verify these are intact and secure.
Administration Set Preparation
To avoid the possibility of fluids contacting the air sensor and pump
mechanism, the solution container should be spiked and the set primed before
opening a pump door. The container should also be placed low enough to
permit a loop in the tubing to the side of the pump. Position the container so
that accidental spillage onto the pump is avoided. Do not stretch the set when
removing it from the packaging.
1.
Using a solution filled container (distilled/sterile water, etc.) and a
Baxter Continu-Flo administration set with at least one Y-site, prepare
the administration set according to instructions accompanying the set.
2.
Splice a tubing segment (part no. 03-02-92-479) into the middle of the
administration set.
! WARNING !
If the pump fails to perform as described or stops with a failure code alarm during
any of the tests herein, troubleshoot it in accordance with Chapter 7,
“Troubleshooting” and repair in accordance with Chapter 8, “Removal/
Replacement Procedures” before placing the pump in service.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...