FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
10-11
Operational Checkout
Battery Check
Battery Check
1.
Verify battery charging and capacity.
2.
Plug the battery into AC power for at least 16 uninterrupted hours before
proceeding to the next step.
3.
Load a primed Baxter “s” suffix standard administration set into the
pump. Set the pump’s PRI RATE to 1901 mL/hr and PRI VTBI to 4800
mL and start both pumps.
4.
Unplug the device while it is running. Verify that the battery icon is lit
and that no change in pumping occurs.
5.
Operate the device on battery power for 2.5 hours. Make sure that the
device does not initiate a LOW BATTERY alert before the KVO alert
occurs.
•
If the unit fails this test, change the battery per “Replacement
of Battery” on page 8-24.
•
If the device operates satisfactorily on battery power, plug it
back in. Recharge the battery by allowing the device to remain
connected to AC power for a minimum of 16 hours.
Panel Lock Test
1.
With the pump running without an alert, press the PANEL LOCK
pushbutton on the rear of the pump.
2.
Verify that the word Loc is displayed in the Main Display.
3.
Verify that the input from front panel keys is not accepted except from
BACKLIGHT
and
TOT VOL/STATUS
.
4.
Press PANEL LOCK again. The Loc message should disappear, and
input from all front panel keys should be accepted.
Free Flow Prevention Test
1.
Prime and load a Baxter "s" suffix standard administration set.
2.
Place the distal end of the set into a container of known weight. Open the
roller clamp and close the pump door and observe the distal end of the
set for 60 seconds after closing the door.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...