07-19-B1-729
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
10-9
Operational Checkout
FSR Check
5.1
Clean the FSR connectors with a nonlubricating contact
cleaner. Access the connectors after opening the pump as
outlined in “Separation of Front and Rear Housings” on page
8-5.
5.2
Perform this procedure again to verify FSR operation. If they
fail again, replace the FSR pair in accordance with
“Replacement of Force Sensing Resistor (FSR) Devices for
Tube Misloading Sensors” on page 8-15.
6.
Open the Pump 1 door and misload the tubing over the left FSR.
7.
Close the door.
8.
Verify that the message display shows CHECK SET LOADING, the
ALARM LED blinks on and off, and the audible alarm tone is sounding.
9.
Record the result on the data sheet.
10.
Open the Pump 1 door and misload the tubing over the right FSR.
11.
Close the door.
12.
Verify that the message display shows CHECK SET LOADING, the
ALARM LED blinks on and off, and the audible alarm tone is sounding.
13.
Record the result on the data sheet.
14.
Open the Pump 2 door and misload the tubing over the left FSR.
15.
Close the door.
16.
Verify that the message display shows CHECK SET LOADING, the
ALARM LED blinks on and off, and the audible alarm tone is sounding.
17.
Record the result on the data sheet.
18.
Open the Pump 2 door and misload the tubing over the right FSR.
19.
Close the door.
20.
Verify that the message display shows CHECK SET LOADING, the
ALARM LED blinks on and off, and the audible alarm tone is sounding.
21.
Record the result on the data sheet.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...