FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
Pump Serial Number:
_____________________________
Printed Name
Signature
Date
Tester:
_____________________________
_____________________________
____________
Reviewer (optional):
_____________________________
_____________________________
____________
N/A _________
Calibrated Syringe Due Date/ID: ________________________________
Door Open and Alarm Volume Test
Pump 1
"DOOR OPEN" message displays
Alarm LED flashes
Audible alarm sounds
No
R
Yes
R
Pump 2
"DOOR OPEN" message displays
Alarm LED flashes
Audible alarm sounds
No
R
Yes
R
Slide Clamp Mechanism Test
Pump 1
"INSERT SLIDECLAMP" message displays
No
R
Yes
R
Pump 2
"INSERT SLIDECLAMP" message displays
No
R
Yes
R
Air Alarm Test
Pump 1
NORM
(50
μ
L) No air detected
Fail
R
Pass
R
(110
μ
L) Air detected
Fail
R
Pass
R
MIN
(25
μ
L) No air detected
Fail
R
Pass
R
(85
μ
L) Air detected
Fail
R
Pass
R
Pump 2
NORM
(50
μ
L) No air detected
Fail
R
Pass
R
(110
μ
L) Air detected
Fail
R
Pass
R
MIN
(25
μ
L) No air detected
Fail
R
Pass
R
(85
μ
L) Air detected
Fail
R
Pass
R
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...