FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
8-5
Removal/Replacement Procedures
Disassembly/Reassembly
Separation of Front and Rear Housings
1.
Turn the pump off and place it face down on an anti-static mat or work
surface. Take care not to lay the pump face down on components (such
as screws) which could damage the front panel.
2.
Remove 10 screws that secure the rear housing to the front housing.
(Figure 11-4 , items 6 and 40.)
3.
Stand the pump upright and separate the front and rear housings by
slowly pulling the front half forward by about 10.16 cm (4”).
4.
While the pump is standing, disconnect CN301 and CN302 from the
CPU board (Figure 11-2 , item 2). The connectors are located at the top,
in the front half of the pump.
Note:
The backup audible alarm may be activated. Continue with the following step
to disable the alarm.
5.
Lay the front case down on its front surface. Position it immediately in
front of the rear housing.
6.
Disconnect the following connectors so that the halves can be
completely separated:
6.1
CN751 from the backup audible alarm board (Figure 11-3 ,
item 2) located at the top left of the rear housing to disable the
backup audible alarm during disassembly.
6.2
Unlock the flat cable connector flange before disconnecting the
cable by first pulling the flange forward about 2 mm (5/64”).
6.3
Disconnect the flat cable connector CNI/F from the CPU board.
6.4
Disconnect CN1 from the power supply board.
6.5
Disconnect the ground wire from the pump head mounting
plate for Pump 2 (Figure 11-2 , item 6).
7.
Reassemble the housings in the following order:
7.1
Connect the ground wire to the pump head mounting plate for
Pump 2.
7.2
Make sure that the connector flange is unlocked and reconnect
the ribbon cable connector CNI/F on the CPU PCB. Push in the
flange to hold the ribbon cable in place.
7.3
Reconnect CN1 on the power supply board.
CAUTION
Avoid stressing the pump’s internal cabling when separating the front and rear
housings.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...