8
Disassembly/Reassembly
8-32
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
LCD Cushion Installation
Note:
These instruction are for pumps that exhibit blanking displays. Cushions are
installed on the inside of the case to keep the case interior from touching the
LCD.
Equipment:
Q
#2 Phillips Screwdriver or power screwdriver
Q
Indelible ink (permanent marker)
Parts:
Q
1 ea. LCD cushion (part no. PCUSU101.A)
Q
1 ea. Upgrade Label (part no. 072601047)
1.
Turn off the pump and make sure it is disconnected from AC power.
2.
Separate the front and rear housings (“Separation of Front and Rear
Housings” on page 8-5).
Note:
Carefully lift and remove the pump head wire assemblies on the CPU board if
they are routed between the hybrid circuits, HIC301 and HIC302.
3.
Remove the circuit boards from the case (“Separation of Printed Circuit
Boards” on page 8-7).
Note:
It is not necessary to disconnect CN201 and CN803. the CPU/Display board
assembly can be placed beside Pump 1.
Figure 8-5 LCD Cushion Placement (Inside View)
Note:
The LCD cushion is installed on the inside of the case housing not on the
LCD.