FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
10-5
Operational Checkout
Door Open Alarm Test
Door Open Alarm Test
1.
Load the primed administration set into the Pump 1 pump head and close
the door.
2.
Program a primary rate of 100 mL/hr and a primary VTBI of 10 mL.
3.
Press the
PRI START
key.
4.
Open the pump door. The following should occur:
•
DOOR OPEN is displayed in the Message Display.
(CLOSECLAMP may also appear depending on the
configuration.)
•
The red ALARM LED flashes.
•
The audible alarm is activated.
5.
Rotate the volume knob through its entire travel. Verify the alarm
volume increases with clockwise rotation, and that the alarm remains
audible at all settings. Leave the volume set to maximum.
6.
Repeat this procedure for Pump 2.
Slide Clamp Mechanism Test
The slide clamp feature is an alert or alarm (software versions 1.09 or later)
option enabled through the configuration utility. When this option is enabled,
the slide clamp on the administration set must be loaded into the slide clamp
slot. When the infusion is complete, the administration set cannot be easily
removed from the pump without first pushing the slide clamp all the way into
the slide clamp slot, thus occluding the tubing in the slide clamp. This test
verifies that the slide clamp feature is functioning properly.
Note:
An administration set with the dark blue slide clamp (part no. 030801011)
compatible with the slide clamp mechanism must be used for this test.
Note:
If the slide clamp feature is enabled, the administration set slide clamp must
be loaded into the assembly. In the alert mode, the pump operates with an
alert if the slide clamp is not loaded. In the alarm mode (software versions
1.09 or later), the pump does not start and an alarm tone sounds if the slide
clamp is not loaded. Use of the slide clamp feature is selected as part of the
configuration option before the pump is placed into service.
Note:
This test is to be performed with the SLIDE CLAMP configuration set to option
2 (ON or ALERT) or 3 (ALARM). If necessary, remove the spring retainer and
enter the modify configuration mode to configure INS CLAMP to option 2 (ON
or ALERT) or 3 (ALARM).
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...