07-19-B1-729
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
9-7
Calibration
Air Sensor Calibration
4.1
Press and hold the
CLEAR TOT VOL
and
2
keys, then press the
ON-OFF/CHARGE
key.
4.2
Release the
ON-OFF/CHARGE
key without releasing the
CLEAR TOT VOL
and
2
keys for at least three seconds.
•
The pump is in Test Mode 2 if NORM and MIN are displayed
on the left hand side of the message (upper) display.
•
If the pump is not in Test Mode 2, turn off the pump and repeat
this step.
5.
Using the set created in step 2, load the fluid filled calibration tubing
segment, into the pump head and close the pump door. Open and close
the pump door two more times and wait 2 minutes.
6.
Adjust the NORM value (PRI RATE window) and the MIN value (PRI
VTBI window) using VR801 for Pump 1 of VR802 for Pump 2. Take the
greater of the NORM and MIN values and calibrate to 450 ±20 (430 -
470).
Q
If the lesser value is less than 410, adjust it up so that it is just above
410. Recheck that the greater of the NORM and MIN values is still
within 430 - 470.
Q
If both the NORM and MIN air sensor values are within the specified
ranges, continue to the next step.
Q
If these ranges cannot be achieved, replace the air sensor per “Air
Sensor Assembly Replacement” on page 8-14. After the air sensor is
replaced, return to step
3 above.
7.
Remove the fluid filled tubing from the pump head.
8.
Load an air filled piece of calibration tubing into the pump head and
close the pump door.
9.
Open and close the pump door two more times and wait 2 minutes.
Q
If the values are 5 or less proceed to the next step.
Q
If the values are greater than 5 perform the following procedure:
9.1
Adjust VR801 and VR802 so that the higher of the NORM or
MIN values is 4 ±1.
9.2
Load the fluid filled piece of calibration tubing into the pump
head and close the pump door.
9.3
Open and close the pump door two more times and wait 2
minutes.
9.4
Verify that both NORM and MIN values are within the range of
410 and 470.
Note:
Apply Red Glpt to VR801 and VR802.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...