07-19-B1-729
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
2-1
Chapter 2
Chapter 2
Description
Overview
This chapter provides an overview of the pump controls and features. This
chapter contains the following information:
Q
“Technical Specifications,” 2-1
Q
“Controls and Indicators,” 2-2
For detailed instructions on using the pump, refer to the pump’s Operator’s
Manual.
Technical Specifications
Item
Characteristic
Catalog Code Number
2M8064
Description
Dual channel linear peristaltic volumetric infusion pump
Administration Set
Baxter’s standard administration set with “s” suffix
Keep Vein Open (KVO) rate
5 mL/hr or programmed rate, whichever is less
Battery
12 Volt, 3.2 Ah sealed lead acid
Battery Life
• Approximately 6 hours with one pump running at a rate from
1 to 1400 mL/hr using a fully charged battery
• Approximately 4 hours with both pumps running at a rate from
1 to 1400 mL/hr using a fully charged battery
Battery Recharge
8 hours for 80% recharge
16 to 72 hours for full charge, depending upon battery condition
AC Power Requirements
110/120V, 60 Hz