FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
7-27
Troubleshooting
Troubleshooting Chart
Inaccurate delivery: Accuracy
Check administration set for proper
type and code.
Check position and seating of the
tubing. Check that ambient and
solution temperatures are between 60°
and 100° F (15.5°and 37.7° C).
Check for cracks in the door hinges.
Check for deformation of the coil
springs of the back plate. Press back
plate in and out and ensure range of
motion.
Check for worn cams and pumping
fingers.
Remove front cover. Check for
damaged or deformed door latch or
damaged latch area.
Replace with Baxter "s" suffix
standard administration set if required.
Position tubing properly, without
stretching or slack.
Raise or lower the temperatures.
Replace the pump head in accordance
with “Replacement of Pump Head
Assembly” on page 8-12.
Replace the springs if required.
Replace worn parts as required.
Replace door latch in accordance with
“Replacement of Pump Door Latch”
on page 8-10.
Replace door in accordance with
“Replacement of the Pump Door or
Pump Door Assembly” on page 8-11.
Operational Issues
A FLOW RATE or CHECK VTBI
alarm occurs when a START key is
pressed.
Check that the rate or VTBI are within
the limits set by the configuration
option.
Change the maximum rate and/or
maximum VTBI setting through the
configuration option.
All front panel keys other than
BACKLIGHT
and
TOT VOL/
STATUS
are disabled.
Does Loc appear in the Main Display?
Press and release PANEL LOCK
switch on the rear panel. Loc in the
Main Display should disappear.
One or more front panel keys are not
accepted.
Check if the front panel is locked (Loc
appears in Main Display).
Turn pump off and back on and try the
key again.
Check front panel ribbon cable for
proper connection and continuity.
Press the PANEL LOCK switch to
remove the panel lock.
Replace front panel film.
Replace Sensor board.
Replace Display board.
The interval between audible alarm
tones is too long.
N/A
Change the interval for alert and/or
alarm tones to the desired value
through the configuration option.
Table 7-4
Troubleshooting Chart — continued
Symptom
Checks
Corrective Action
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...