A
A-4
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
Pump Serial Number:
_____________________________
Printed Name
Signature
Date
Tester:
_____________________________
_____________________________
____________
Reviewer (optional):
_____________________________
_____________________________
____________
Multimeter Calibration Due Date/ID: _________________________________
Electrical Safety Tester Calibration Due Date/ID: _______________________________
Battery Check
Pump Runs 2.5 hrs. Without Battery Low Alert
Record Run Time:
Panel Lock Test
"Loc" displays, front panel keys (except
BACKLIGHT
and
TOT/VOL/STATUS
) do not
accept input
Fail
R
Pass
R
Free Flow Prevention Test
Pump 1
Pump Prevents Fee Flow of Fluid
< 0.5 mL
mL
Pump 2
Pump Prevents Fee Flow of Fluid
< 0.5 mL
mL
Electrical Safety Tests
Ground Impedance
Ground Impedance
< 0.5
Ω
Ω
Leakage Current
Earth Leakage Current
< 50
μ
A
μ
A