6
Cleaning
6-2
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
For a pump that has been in an Isolation Area, select those agents from Table
6-1 that both clean and disinfect. Only external parts of the pump should be
disinfected. The following are procedures for cleaning accessible areas of the
pump. Do not use hard instruments for cleaning.
1.
Lift the door latch to the open position. Open the door and press the
safety clamp latch until it locks in the open position.
2.
Using a cotton swab dampened with one of the agents listed in Table 6-1,
clean all tubing guides and tubing channels from the top of the pump to
the exit point below the safety clamp.
3.
Clean all surfaces in the pump head which may contact the tubing.
4.
Clean all surfaces of the air sensor located just above the safety clamp.
This area must be completely dry and free of foreign matter prior to
reuse.
CAUTION
CAUTION
Table 6-1
Recommended Cleaning Solutions
Cleaner
Septisol
Super Edisonite
Bafix
Tor
Hi-Tor Plus
10% bleach and water
Soapy water
Isopropyl alcohol (up to 95%)