First installation
59
Measure Now
Click on this button if you wish to measure the actual luminance value.
You will be asked to calibrate the sensor (if necessary) and apply it to the
screen.
Display function
Note: This item is grayed for predefined Presets for Barco projectors (in
Dicom Theater). If you create a new Preset, you will be able to define a
display function.
Select the required display function from the drop-down list.
Select the required display function from the drop-down list. For most
medical viewing applications, we recommend you select the DICOM
display function, as this function results in the most visible
grayscales.DICOM display function, as this function results in the most
visible grayscales.
If you select Gamma, you can enter a display gamma value if the imaging
board head, to which the display is connected, has an editable gamma
table.
If you select Uncorrected, the program fills the imaging board gamma
table with a linear function.
Ambient light compensation
Check this option if you want to compensate for the influence of the
ambient light on the image. Especially when you have selected a DICOM
or CIE display function, we advise to switch on this compensation to
obtain a better display performance.
Glare compensation
Check Glare Compensation if you want to compensate for the influence of
the bright parts on the dark parts of the image.
This option is always on (and grayed) if a DICOM display function is
selected.
2.8
Calibrating displays
After defining the display’s Preset, the display is calibrated. This is
described in detail in § “Calibrating displays” on page 59.
2.9
Setting up the Q/A schedule
When every display in the configuration has the correct Preset values and
is calibrated successfully, you need to set up the schedule of Q/A tasks
that is appropriate for this viewing station.
Summary of Contents for MediCal Pro
Page 1: ...Installation User Manual MediCal Pro...
Page 2: ...www barco com B4100132 02 July 2004...
Page 9: ...Introduction 9 INTRODUCTION...
Page 18: ...Introduction 18...
Page 19: ...First installation 19 FIRST INSTALLATION...
Page 65: ...65 USING MEDICAL PRO...
Page 104: ...104 About Displays the details of this version of MediCal Pro...
Page 105: ...105...
Page 106: ...106 GLOSSARY...
Page 110: ...110...