First installation
58
On color displays and projectors
Measure on next calibration:
Select this option if you do not wish to enter a target value now. Instead,
the target value will be determined by measuring during the next
calibration.
Absolute value
Select this option if you wish to enter a numeric target value.
This option is preferred for Barco displays.
Luminance
In this field, enter the required target value, expressed in Cd/m².
The field will be grayed if the “Measure on next calibration” option is
selected.
Chroma (color displays only)
Note: This item is grayed for predefined Presets for Barco projectors (in
Dicom Theater). If you create a new Preset, you will be able to define a
chroma.
In this field, enter the required color temperature or chroma value,
expressed in Kelvin or CIE (x,y) units.
In the Black luminance option window, you can check the option Same as
white chroma in case you wish the black and white chroma values to be
the same.
Summary of Contents for MediCal Pro
Page 1: ...Installation User Manual MediCal Pro...
Page 2: ...www barco com B4100132 02 July 2004...
Page 9: ...Introduction 9 INTRODUCTION...
Page 18: ...Introduction 18...
Page 19: ...First installation 19 FIRST INSTALLATION...
Page 65: ...65 USING MEDICAL PRO...
Page 104: ...104 About Displays the details of this version of MediCal Pro...
Page 105: ...105...
Page 106: ...106 GLOSSARY...
Page 110: ...110...