2 Put the sensor without Ambient Light Shield (for convenience) on the
image and follow the instructions on the screen.
The full screen calibration starts with a measurement of luminance in the
center of the screen. It proceeds with calibration in a number of zones
over the picture tube so that the luminance in all of the zones equals the
center luminance.
The zone where you have to put the sensor, is indicated by a white,
blinking square.
Calibrating non-Barco displays
1 If you use an X-Rite DTP92 sensor, you may be asked to calibrate the
sensor first.
Make sure the sensor is completely covered (so that no light can enter
the sensor) and click on OK.
2 Put the sensor, with Ambient Light Shield if appropriate, on the image
on the screen.
Important:
Make sure there are no application windows or other items (e.g., the
cursor) displayed under the sensor. They would cause the calibration
to fail.
When ready, click on OK.