Clicking on the Contrast (Brightness) icon left of the sliders, puts the
Contrast (Brightness) control in calibrated position. This is the preferred
position.
Luminance Uniformity Correction on/off
This function is available for BARCO displays only.
In the Configuration section in the main window, right-click on the icon of
the display of which you wish to switch LUC on or off.
From the drop-down menu that appears, select Luminance Uniformity
Correction...
The Luminance Uniformity Correction system, if calibrated by means of
“Full Screen Calibration” (see below), ensures a uniform luminance, even
in the corners of the picture tube.
This function is grayed if the function is not available on the display.
Full Screen Calibration
This function starts with a normal calibration in the center of the
screen. It proceeds with calibration in a number of zones over the picture
tube.
Please refer to § Full screen calibration : “ Full screen calibration “ for
more information.
The function is grayed if not available on the display.
Degaussing displays
This function is available for color BARCO displays only.
In the Configuration section in the main window, right-click on the icon of
the display you wish to degauss.
Summary of Contents for MediCal Pro
Page 1: ...Installation User Manual MediCal Pro...
Page 2: ...www barco com B4100132 02 July 2004...
Page 9: ...Introduction 9 INTRODUCTION...
Page 18: ...Introduction 18...
Page 19: ...First installation 19 FIRST INSTALLATION...
Page 65: ...65 USING MEDICAL PRO...
Page 104: ...104 About Displays the details of this version of MediCal Pro...
Page 105: ...105...
Page 106: ...106 GLOSSARY...
Page 110: ...110...