3 For sensors that have an ambient light shielding only: If you have
specified an absolute value for black and white luminance in the
Preset, and the Ambient Light Compensation option is checked, the
software asks you to remove the sensor from the screen and put it
back without Ambient Light Shield.
When ready, click on OK.
After a few measurements, the software asks to remove the sensor
once more, and put it back with the Ambient Light Shield.
When ready, click on OK.
4 The black luminance target value is determined now.
The sensor measures a black patch. The result of the measurement is
shown in a small Measure window in the corner.
a) If you have specified an absolute value for black luminance in the
Preset, MediCal Pro displays this value in the message.
You must control the display brightness manually until the
measured value is equal or lower than the target value.
When this is the case, the OK button becomes active so that you
can continue.
b) If you selected Measure on next calibration in the Preset, you
must adjust display brightness manually, until a desired level of
black luminance is attained.
When you click on OK, the measured value will be taken as black
luminance target.