93
3.10 Selecting Luminance units
The results and values of luminance measurements can be expressed
in Candela per square meter (Cd/m²) or Footlambert (fL).
To select the luminance units:
1 From the Tools menu, select Options > Application Settings (see
above).
2 Select the General tab (see above).
3 Select the desired luminance unit.
3.11 Selecting units of length
In the Display properties, you can enter the screen size of the displays.
You can do this in inches or millimeters. You can select this unit of length
in the Application Settings dialog.
To select the units of length:
1 From the Tools menu, select Options > Application Settings (see
above).
2 Select the General tab (see above).
3 Select the desired unit of length.
3.12 Setting up optical sensor list and priority
MediCal Pro supports a list of possible optical sensors. To speed up the
search for connected optical sensors, it is best to indicate which sensor
types are used and which ones are not used.
Moreover, you can set up a priority of sensors. This may be necessary
when using Barco flat panel displays with built-in I-Guard sensor. When
you wish to use an external sensor for measurements on the flat panel,
you should assign it a higher priority than the I-Guard sensor. In that case,
MediCal Pro will automatically select the external sensor for the
measurements.
To set up the list and priority of used sensors:
1 From the Tools menu, select Options > Application Settings (see
above).
Summary of Contents for MediCal Pro
Page 1: ...Installation User Manual MediCal Pro...
Page 2: ...www barco com B4100132 02 July 2004...
Page 9: ...Introduction 9 INTRODUCTION...
Page 18: ...Introduction 18...
Page 19: ...First installation 19 FIRST INSTALLATION...
Page 65: ...65 USING MEDICAL PRO...
Page 104: ...104 About Displays the details of this version of MediCal Pro...
Page 105: ...105...
Page 106: ...106 GLOSSARY...
Page 110: ...110...