First installation
52
Horizontal geometry settings
Horizontal position (course)
Position the middle vertical line of the test image approximately in the
center. This is a course adjustment. You will get the opportunity to fine-
tune later in the procedure.
Screen width
Adjust the width of the white frame on the test image following the
display’s specifications. You will get the opportunity to fine-tune later in
the procedure. Tip: It is a good idea to use a flexible plastic or paper ruler
to measure the size.
Horizontal linearity
Adjust until the distance between the vertical lines on the left side of the
test image is the same as between those at the right. These lines are
marked by arrows.
Horizontal S-correction
Adjust until the distance between the vertical lines in the center of the
test image is the same as between those at the left and the right.
Horizontal blanking left
The background of the test image is highlighted (gray) and a white frame
appears around the test image. When you drag the slider in the
adjustment dialog to the right, a black vertical border appears at the left
side of the test image. This is the blanking. Adjust now until this blanking
stops at 0.5 cm (0.2 inch) before the left edge of the white frame. The
frame must be completely visible.
Horizontal blanking right
Now do the same at the right side of the test image. Adjust until the
blanking starts at 0.5 cm (0.2 inch) after the right edge of the white
frame . The frame must be completely visible.
Summary of Contents for MediCal Pro
Page 1: ...Installation User Manual MediCal Pro...
Page 2: ...www barco com B4100132 02 July 2004...
Page 9: ...Introduction 9 INTRODUCTION...
Page 18: ...Introduction 18...
Page 19: ...First installation 19 FIRST INSTALLATION...
Page 65: ...65 USING MEDICAL PRO...
Page 104: ...104 About Displays the details of this version of MediCal Pro...
Page 105: ...105...
Page 106: ...106 GLOSSARY...
Page 110: ...110...