![background image](http://html1.mh-extra.com/html/barco/coronis-1mp/coronis-1mp_installation-and-user-manual_3869470083.webp)
BarcoMed Driver Tab
Single Source Master
84
Disable 10-bit grayscale
Use this option for standard Windows®-based applications that
require 8-bit color support (256 colors), such as Internet Explorer,
Excel, etc.
Do not use this option for applications that require 10-bit pixel
support (1024 simultaneous shades of gray) and use extended
depth graphic libraries such as WinBarco.
Description of 10-bit grayscale
To show 1024 simultaneous shades of gray all BarcoMed “H”
series display controllers, such as the BarcoMed 1MP2FH, 2MP2H,
5MP2FH or BarcoMed Coronis for example, run in extended depth
(10-bit) mode by default.
Implications
When turned on the “Disable Extended Depth” option causes a
BarcoMed “H” series display controller to run in 8-bit mode,
which results in increased performance since the frame buffer is
set to run with 8-bit pixels. While this option may be used with
applications that normally require 10-bit support, there will be a
loss of pixel depth, i.e. 256 shades of gray instead of 1024
simultaneous shades of gray.
Monitor Configuration
If the options in this section are grayed out, then they are not
available for the model controller with which you are working.
Under Windows 2000 and Windows XP you can
enable
or
disable
DualView through the drop-down menu.
In SingleView mode the Monitor Configuration section allows you
to select the number of monitors that the current display boards
should drive, as well as how the monitors should be positioned.
For example
, if the current board installation is capable of
driving four heads, but you only have three monitors which are
placed in a single row, then you can select the “Three monitors–
Summary of Contents for CORONIS 1MP
Page 1: ...Installation User Manual Coronis 1MP...
Page 12: ...13 Display Controller Installation...
Page 13: ...Display Controller Installation 14 This page intentionally left blank...
Page 20: ...21 Display Installation...
Page 21: ...Display Installation 22 This page intentionally left blank...
Page 36: ...37 Software Installation...
Page 37: ...Software Installation 38 This page intentionally left blank...
Page 70: ...71 Display Controller Tools...
Page 71: ...Display Controller Tools 72 This page intentionally left blank...
Page 75: ...Barco Display Tab Single Source Master 76 Figure 38...
Page 88: ...BarcoMed Hardware Tab 89 Figure 44 BarcoMed Hardware Tab under Windows NT 4 0...
Page 96: ...97 Display Tools...
Page 97: ...Display Tools 98 This page intentionally left blank...
Page 109: ...Barco DPMS Screen Saver Single Source Master 110 This page intentionally left blank...
Page 113: ...Barco I Switch Software 114 This page intentionally left blank...
Page 114: ...115 Display Operation...
Page 115: ...Display Operation 116 This page intentionally left blank...
Page 148: ...149 MediCal Software Installation and Operation...
Page 149: ...MediCal Software Installation and Operation 150 This page intentionally left blank...
Page 152: ...153 Troubleshooting...
Page 153: ...Troubleshooting 154 This page intentionally left blank...
Page 166: ...Single Source Master Driver re installation updates or removal 167...