Azbil Corporation
Maintenance
Model SLX 110/120 - Smart displacement type level transmitter 3-23
3-6 : Returning all set data to factory settings
This procedure can return not only calibrated data but also all other set data (damping
time constant, etc.) to factory settings.
Before restoring the factory settings, it is recommended to record or print out the
currently set data.
Step
Button
Procedure
SFC screen
1
SHIFT
Press [SHIFT].
SHIFT-
2
COR- K
RECT
RESET
Press [CORRECT].
SLX TAG - 0001
RESET CORRECTS ?
3
ENTER
( Yes )
NON-VOL
Press [ENTER]. A confirmation screen
will appear.
SLX TAG-0001
ARE YOU SURE ?
4
ENTER
( Yes )
NON-VOL
Press [ENTER] again.
SLX TAG - 0001
WORKING...
5
A
ID
DE READ
All set data returns to factory settings.
Press [ID].
SLX TAG - 0001
CORRECTS RESET
PRESS ID
6
CLR
( No )
The current settings (factory settings)
are loaded to the SFC. Press [CLR] to
exit.
SLX TAG - 0001
READY...