Medical electrical equipment needs special precautions regarding EMC (Electromagnetic compatibility) and needs to
be used according to the EMC information below.
The
Century
bath is intended for the use in the electromagnetic environment specified below. The customer or the user
of the
Century
bath should assure that it is used in such an environment.
WARNING
Stacking or placing other electrical equipment next to this device is not recommended, it can
interfere with the equipment’s operation and safety. Portable and mobile radio-frequency (RF)
communications equipment can interfere with this equipment operation and safety.
WARNING
The equipment may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take action, such as reorienting, relocating the equipment
or shielding the location.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Emissions test
Compliance
Electromagnetic Environment – Guidance
RF emissions
CISPR 11
Group 1
The
Century
bath uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
Century
bath is suitable for use in all establishments, those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emission
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emission
IEC 61000-3-3
Complies
Electromagnetic Compatibility