13
Scalding Protection
Century
bath is equipped with an electric scalding protection to
reach a maximum safety for the patient. The function is used
during filling, showering and cleaning.
The scalding protection is activated when the water temperature
gets too high. If the bathtub water reaches 45 °C (113 °F) or
above, the electronic scalding protection shuts off the water after
10 s, both LEDs on the Disinfection/Autoclean button and the
Patient Shower button flash for a short time.
Reset the scalding protection (5 steps)
1
Make sure both LEDs on the Disinfection/Autoclean button
and the Patient Shower button have stopped flashing.
2
Turn the knob on the thermostatic mixer to COLD.
3
Press the Tub Filling button until the display shows 30-
35 °C (86-95 °F).
4
Press the Patient Shower button. Lift off the shower handle,
direct it away from yourself down into the bathtub. Press the
trigger on the shower handle.
5
Allow a few seconds for the water to reach correct
temperature.
Patient Shower
(7 steps)
1
The patient shower is on the right side of the panel.
(See Fig. 4)
2
Press the Patient Shower button on the control panel.
The green LED indicates when the patient shower is
activated.
3
Lift off the shower handle, direct it down into the bathtub.
Press the trigger on the shower handle.
(See Fig. 5)
Continue with the steps on the next page.
WARNING
To prevent scalding if the scalding protection is
activated, remove the patient from the bath
immediately in a safe manner.
NOTE
The shower handle and the fill function cannot be
activate at the same time.