PROBLEM
ACTION
Hydromassage/Hydrosound function does not start.
•
Check power supply and fuses.
•
Make sure that no other function is activated.
•
Make sure that the water level is above the top
level sensor in the bathtub.
•
Check the
Hydrosound
system LED, a yellow flash
indicates that a problem has occurred.
•
Contact qualified personnel.
Both LEDs at the Disinfection/Autoclean button and the
Patient Shower button are flashing.
•
Scalding protection is activated. See
Scalding
Protection on page 13
.
•
Contact qualified personnel.
WARNING
To prevent scalding, always check the
water temperature with your naked hand
before directing the water on the patient.
Do note use gloves as it may insulate to
the extent that the water temperature can
be misjudged. Point the flow of the water
away from the patient.