27
40 Wet a new cloth with water and remove all traces of
disinfectant on contact areas. The cloth should occasionally
be rinsed in running water when removing disinfectant.
41 Leave to air dry.
42 Attach any accessories.
Clean and Disinfect Hydromassage™ System
with Automatic Disinfection
(15 steps)
This procedure is a complement when cleaning the bath system,
see section
Clean and Disinfect with Disinfection Spray Handle
or Spray Bottle on page 24
.
To ensure proper disinfection, the
Hydromassage
system is
automatically disinfected. The treatment hose is an accessory. If
you do not have a treatment hose, skip such cleaning instructions.
1
Check the level of disinfectant.
2
Open the drain.
3
Remove the suction inlet screen and clean it separately.
(See Fig. 1)
4
For treatment hose only: pull out the
Hydromassage
system
nozzle and clean it separately.
(See Fig. 2)
5
For treatment hose only: Insert the treatment hose in the
Hydromassage
system outlet.
(See Fig. 3)
6
Press the Disinfection button on the control panel to start the
automatic disinfection cycle. Disinfectant is flushed
through the
Hydromassage
system nozzle/treatment hose.
The disinfection cycle ends with the
Hydromassage
system
being rinsed with water.
7
For treatment hose only: When the disinfectant cycle is
completed, remove the treatment hose and let it hang dry on
the holder. Place the
Hydromassage
system nozzle back into
the
Hydromassage
system outlet.
Continue with the steps on the next page.
NOTE
Let the pillow dry with the suction cups upwards.
NOTE
Look at the flow meter and make sure there is an
appropriate mixture of disinfectant liquid to the water,
according to section
Replace Disinfectant Bottle on
page 23
.
Fig. 1
Fig. 2
Fig. 3