Control Panel Problems
PROBLEM
ACTION
No shampoo, body wash, bath oil/
ArjoSound
water
conditioner emerges from the dispenser when pressing
the Shampoo & Body Wash or Bath Oil/
ArjoSound
water conditioner button.
•
Check power supply and fuses.
•
Check level in Shampoo/Bath oil container.
•
Contact qualified personnel.
Main power failure.
•
Check power supply and fuses.
•
Contact qualified personnel.
Temperature display read-out failure.
•
Check power supply and fuses.
•
Contact qualified personnel.
No reaction when pressing a button on control panel.
•
Check power supply and fuses.
•
Make sure that no other function is activated.
•
If the shower and disinfection LED are flashing,
the scalding protection is activated, see
Scalding
Protection on page 13
.
•
Contact qualified personnel.
Disinfection does not start when pressing Disinfection
button.
•
Check power supply and fuses.
•
Make sure the bathtub is empty of water.
•
Make sure shower, fill or
Hydrosound
system is not
activated.
•
Check level in the disinfection container, replace if
low.
Replace Disinfectant Bottle on page 23.
•
Contact qualified personnel.
WARNING
To prevent scalding, always check the
water temperature with your naked hand
before directing the water on the patient.
Do note use gloves as it may insulate to
the extent that the water temperature can
be misjudged. Point the flow of the water
away from the patient.