16 Spray disinfectant into the surface overflow outlet.
(See Fig. 4)
17 For disinfection spray handle only: Press the Disinfectant
button on the control panel to turn it off, and place the
disinfectant spray handle back behind the access door.
Scrub the surface of the bathtub, loose parts and accessories
with a soft bristle brush or cloth (to remove any deposits e.g.
skin flakes). Scrub thoroughly, especially on handles and
hand control.
(See Fig. 5)
19 Open the drain.
20 Press the Patient Shower button on the control panel.
21 Rinse all parts clean with plenty of water having a
temperature of approximately 25 °C (77 °F), using the
shower handle, to remove disinfectant.
22 Soak a cloth in disinfectant.
23 Wipe all other contact areas with the cloth, such as the
outside of the bath (particularly behind the head), sensors
and hand control.
24 Wet a new cloth with water and remove all traces of
disinfectant on contact areas. The cloth should occasionally
be rinsed in running water.
25 Examine bathtub and accessories to make sure that all of the
disinfectant is removed. Repeat rinse if needed.
26 Examine bathtub and accessories for visible damage.
For Century bath with Hydromassage system, go to section
Clean and Disinfect Hydromassage™ System with Automatic
Disinfection on page 27, after that proceed with Disinfect (Step
27-42 of 42) on page 26.
Continue with the steps on the next page.
18
CAUTION
Never use green or brown Scotch-Brite nor steel
wool on the bath as it may scratch the surface.
Fig. 4
Fig. 5