Technical Data: Still Bath, Hydrosound® System & Hydromassage™ System
NOTE
All drainage, water and electrical installation work should be carried out by qualified personnel, in
accordance with local and national regulations.
Market
Voltage
Frequency
Fuse (slow blow)
Power
consumption
Europe
230/240 V single
phase AC
50 Hz
6 A
maximum
700 VA
North America
120 V single
phase AC
60 Hz
10 A
maximum
700 VA
Japan
100 V single
phase AC
50/60 Hz
10 A
maximum
700 VA
General
Power supply cord: replaced by qualified personnel
The unit must be continuously powered and connected to a separate fuse with a ground fault circuit
interrupter.
Fuses according to EN60898.
Main switch must be placed close to the bathtub and be visible
for the user.
Power supply is activated via the main switch.
The unit must be permanently connected to the mains when used in USA and Canada. All electrical
installations must comply with the US National Electrical Code and the Canadian Electrical Code
respectively.
Pollution degree
2
Thermometer
Range
10-50 °C (50-122 °F)
Accuracy
+/-1 °C (1.8 °F)
Resolution
0.1 °C (0.18 °F)
Loading
SWL (Safe Working Load, p water)
380 kg (838 lbs)
Maximum patient weight
182 kg (401 lbs)
Maximum floor loading/foot pad
482 kg / 0.4 m
2
Maximum floor pressure
0.5 N/mm
2
Water Drainage
Minimum drainage capacity
100 l/minute
Technical Specifications