14
Direct the water stream away from the patient and allow a
few seconds for the water to reach desired temperature. If
desired, adjust the water temperature by using the knob on
the panel.
5
Place your hand in the water stream to make sure the water
temperature feels comfortable before the water stream is
directed on the patient.
To get continuous water flow without having to hold
in the trigger:
6
Press the shower handle trigger and push in the Lock button
on the side. The water will continue to flow without having
to press the trigger.
(See Fig. 6)
7
Release the lock by pressing the handle trigger and the Lock
button will pop out by itself.
Disinfection Unit
The bath has a built-in disinfection unit. The disinfection bottles
are placed behind a door on the left side of the panel together with
a disinfection spray handle and a flow meter. The flow meter
shows the liquid mix ratio between disinfectant and water. The
disinfectant bottle cap is adjustable and regulates the
disinfectant/water ratio.
(See Fig. 7)
For further cleaning and disinfection instructions, see section
Cleaning and Disinfection Instructions on page 21
.
4
WARNING
To prevent scalding, always check the water
temperature with your naked hand before
directing the water on the patient. Do note use
gloves as it may insulate to the extent that the
water temperature can be misjudged. Point the
flow of the water away from the patient.
Fig. 6
Lock Button
Trigger
Fig. 7