8.3.2.E
Perform Verification
Available on the MAX
®
4000XL only
To get here, navigate:
[MENU]
CALIBRATION MENU
TEMPERATURE CALIBRATION MENU
PERFORM VERIFICATION
This selection performs the temperature
verification of the MAX
®
4000 when the
Temperature Calibration Kit is connected.
Performing the verification:
•
Connect the Temperature Calibration
Kit as described in section
•
Select the PERFORM
VERIFICATION menu item.
•
The introduction screen is displayed;
Follow the instructions.
•
Then press
[GO]
to continue.
•
During the verification, press
[DETAIL]
to view real-time results at
each verification point, if desired.
•
Wait for the verification to finish.
•
Open the lid and wait for the MAX
®
4000XL to cool.
•
Disconnect the TCI, and remove the
temperature calibration disk.
•
Gently but firmly install the pan
support into the balance stem.
During the verification, the Temperature Calibration Interface (TCI) Module reports its
temperature to the MAX
®
4000XL. The verification process monitors the MAX
®
4000XL’s
temperature (in the heat chamber) and the TCI’s temperature, and reports the difference
between the two.
The verification can be run at any time, and can also be set to run automatically after a
temperature calibration, as set in section